6.3.11 Protection fuses
Protection fuses are foreseen in the following
circuit:
•
Main power supply
•
Socket for auxiliary devices (max. 6 A)
•
12Vdc light led power supply
•
5Vdc power supply for valves group heating
•
12Vdc flowmeter box power supply
•
12Vdc ventilator power supply
•
12Vdc battery
WARNING! Risk of injury for the user / patient
The operations described here below must be carried out by highly
qualified technical personnel, trained and formally authorised by SIARE
only.
In case of break of protection fuse:
•
interrupt the mai power supply;
•
eliminate the abnormal or the cause which caused the
fuse break,
•
replace the protection fuse with one with the same value
and technical characteristics.
WARNING! Risk of injury for the user / patient
Fuses with wrong values and technical characteristics can compromise
equipment integrity and safety.
MORPHEUS
6-19