7.2.3 Electronic flowmeter box pre-setting

After calculation of Minute Volume needed by the patient basing physiologic and
pathologic conditions:
obtained from the product of Tidal Volume x Respiratory Rate (Vti x RR)
Determinate the total flow of fresh gas to be introduced in the patient circuit basing
on chosen dosage.
Press the membrane key to select the
anaesthetic gas to be combined with the
oxygen (Air or N 2 O).
Ad just the flow by the regulator.
7-6
CAUTION. Risk of injury for the patient.
The two gas (Air or N 2 O) cannot be used at the same time.
The opening of N 2 O regulator determines the automatic opening of
O 2 in the measure of 20-30% (MIX-LIFE).
User's Manual, version DU3300105