11.3 Electromagnetic compatibility tables

11.3.1 Annex A: Table 1

Guidance and manufacturer's declaration – electromagnetic emissions
The MORPHEUS is intended for use in the electromagnetic environment specified below. The customer or
the user of the MORPHEUS should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
MORPHEUS
Compliance
Virdict
Group 1
Compliance
Class A
Compliance
Class A, B, C, D, o
Class A
NOT APPLICABLE
Compliance
Compliance
Electromagnetic environment – guidance
The MORPHEUS uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The MORPHEUS is suitable for use in all
establishments other than domestic, and
may be used in domestic establishments
and those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes, provided the following warning
is heeded:
Warning:
This equipment/system is
intended for use by healthcare
professionals only. This equipment/
system may cause radio interference or
may disrupt the operation of nearby
equipment. It may be necessary to take
mitigation measures, such as re-orienting
or relocating the [ME EQUIPMENT or ME
SYSTEM] or shielding the location.
11-15