PAW low
Rate high
9-6
Check that the luminous PAW bar on ventilator (the airways
pressure curve) correctly follows the inspiration / expiration cycle.
Check that nothing is limiting the patient's respiratory capacity.
If this is not the case, contact the Siare Service Centre or a Centre
authorised by Siare.
In this condition, the patient circuit + patient system presents a
lower resistance than expected or a higher compliance. This
causes insufficient ventilation pressure.
Check that the corresponding alarm limits are set correctly.
Check that the mask, endotracheal tube and patient circuit are not
in some way split, disconnected or connected wrongly If this is the
case, eliminate the problem or replace them.
Check the correct settings of the patient's respiratory parameters
(according to the operative mode selected: Volume/Flow, Rate,
I/E, Trigger).
Check that the luminous PAW bar on ventilator (the airways
pressure curve) correctly follows the inspiration / expiration cycle.
Check that the patient circuit is connected correctly to the
anaesthesia unit and to the patient.
Check that the anaesthesia unit delivers the gas mixture correctly.
Check that the low pressure level is higher than the PEEP level
set. If not, increase it above the PEEP level.
If this is not the case, contact the Siare Service Centre or a Centre
authorised by Siare.
This alarm is activated when the breathing rate volume is higher
than the set value.
Check that the corresponding alarm limits are set correctly.
Check that the patient's respiratory parameters are set correctly.
Check that the sensitivity of the Trigger is appropriate to the
patient's physiological conditions.
Contact the Siare Service Centre or a Centre authorised by Siare
User's Manual, version DU3300105