Siare Morpheus E User Manual page 241

Anaesthesia unit
Table of Contents

Advertisement

Low pressure
High pressure
Low frequency
High frequency
Min. Exp. Volume
Max. Volume Exp.
FiO 2 Min.
FiO 2 Max.
Power failure
(electric power
supply)
8-20
This entry is used to set the low pressure alarm limit
02 – 20 cmH 2 O
This entry is used to set the high pressure alarm limit
20 – 80 cmH 2 O
This entry is used to set the low respiratory frequency alarm limit
01 – 11 bpm
This entry is used to set the high respiratory frequency alarm limit
13 – 130 bpm
This entry is used to set the alarm limit for the low expiratory tidal
volume
0 – 1400 ml
This entry is used to set the alarm limit for the high expiratory tidal
volume
1 – 1700 ml
This entry is used to set the limit for the low FiO
20 – 98 %
This entry is used to set the limit for the high FiO
21 – 99 %
This entry permits to enable or disable the acoustic signal of power
failure
Enabled / Disabled
%
2
%:
2
User's Manual, version DU3300105