6.3

Preparation to use

6.3.1 Medical gas connection

Screw the gas supply hoses (O 2 , AIR, N 2 O) to
relevant connectors of the gas supply group.
Screw the gas supply hoses (O 2 , AIR, N 2 O) to
the relevant connectors of the hospital
distribution system (O 2 , AIR, N 2 O).
Ensure that all the supply hoses are connected
and are correctly working.
6-8
The compressed air must be of medical type, therefore oil free
and filtered.
The equipment can also work with oxygen supply only, but in this
case the FiO
will be adjustable at 99% only. An alarm on
2
flowmeter, indicats low air condition will occur at start-up of the
equipment.
The O 2 , N 2 O, AIR hoses are already supplied with screw
connections DISS type (Diameter Index Safety System) for
connection to the anaesthesia unit connectors.
The installation technician taking care of the hoses connection (O 2 ,
N 2 O, AIR) to the outlets must ensure that they are compatible with
the hospital medical gas pipeline system.
To prevent inversions of gases that can be FATAL for the patient,
assembly of the connectors compatible with the hospital distribution
system and all the maintenance and/or replacement operations of
the medical gas supply hoses must be carried out by highly qualified
technical personnel only.
WARNING! Risk of equipment failure.
In order that the anesthesia machine operates as specified, the inlet
pressures of medical gases must be between 280 kPa and 600 kPa
(2.8 - 6 bar).
Before use ensure that this requirement is met.
User's Manual, version DU3300105