Siare Morpheus E User Manual page 269

Anaesthesia unit
Table of Contents

Advertisement

Frequency
Several times a day
Every day / when
necessary
According to local
practice and
standards
Every week / when
necessary
Every 6 months or
1000 working hours
Every 6 months or
1000 working hours
10-10
Component
Patient circuit
Condensation trap
filter
Filters
Anaesthesia unit
Breathing system
Flow sensor
Breathing system
Oxygen sensor
Anaesthesia unit
Oxygen sensor
Filters
Patient circuit
Washers / O-Rings
Procedure / Action
Check for any water collection, drain and clean
the tubes when necessary.
Check for wear.
General cleaning and checks.
Dismount the components and clean, then
sterilize in autoclave, disinfect by steam or
chemically.
Disinfect with steam or chemically.
Dismount the components and clean, then
sterilize in autoclave, disinfect by steam or
chemically.
Calibrate according to the procedures
described in this manual.
The anaesthesia unit must be inspected and
checked in general and any worn parts must be
replaced.
Use the appropriate preventive maintenance kit.
This operation must only be carried out by
qualified technical personnel, according to the
instructions contained in the relative service
and maintenance manual.
Replace.
The working life of the cell depends on the
working environment. If the temperature or the
O2 % is high, the working life of the sensor will
be lower.
Replace.
Sterilize according to the procedure described
in this manual and according to local standards.
Components that cannot be destroyed should
be sterilised before disposal.
User's Manual, version DU3300105