Siare Morpheus E User Manual page 209

Anaesthesia unit
Table of Contents

Advertisement

Pause
PEEP
PLIM
PMax
7-32
INSPIRATORY PAUSE
Through this parameter it is possible to adjust the inspiratory pause timing,
which is measured on display by the % of inspiratory time.; furthermore it is
used for calculation of the respiratory mechanics parameters (compliance
and resistance).
Its duration can be extended from 0% to 60% of inspiratory time with steps
of 1%.
This parameter is used in the following operative modality: VA/VAC.
END EXPIRATION POSITIVE PRESSURE
With this parameter it is possible to set the value of positive airways
pressure in expiratory phase. The PEEP is the positive pressure
maintained in patient circuit during the expiratory phase.
It is possible to ad just this parameter on values from OFF, 3 a 30 cmH
with steps of 1 cmH
O.
2
This parameter is used in all foreseen operative modalities.
The introduction of PEEP during the automatic ventilation requires
caution since it determines the increase of max and mean airways
pressure.
This change requires probably also the modification of alarm limits.
LIMIT PRESSURE
With this parameter it is possible to set the value of max airways pressure.
The parameter is used in the pressure controlled modalities and it is used
to fix an insuperable safety limit on the airways pressure.
It is possible to adjust this parameter on values from 6 to 60 cmH
steps of 1 cmH
O.
2
This parameter is used in the following operative modality: APCV
MAXIMUM PRESSURE
With this parameter it is possible to set, in APCV and VC/VAC BABY
modalities, the max value of the pressure which can be reached to deliver
the set volume.
It is possible to select value from: 6 to 60 cmH
O with
2
O.
2
User's Manual, version DU3300105
O
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Morpheus seriesMorpheus ltMorpheus m