Siare Morpheus E User Manual page 91

Anaesthesia unit
Table of Contents

Advertisement

ENERGY SAVING
BRIGHTNESS
DATE/TIME
APNOEA TIME
GRAPHICS
VISUALIZZATION
5-18
It is possible to set the percentage value of the energy
consumption of the machine. The energy saving condition is
active after 30 minutes calculated from the moment the keys
are not pressed, either the soft keys or the encoder.
If the value 0% is set the screen, when the energy saving
function is activated, will be black.
0 - 100
It is possible to set the brightness of the screen
1 - 30
It is possible to set the data and hour field. Pressing the
encoder it will be possible to select single fields and, by
turning the knob, it will be possible to set and /or modify the
desired date and hour.
Press the STAND-BY soft key to return to SETUP- MAIN
MENU view.
Selecting the "Apnoea Time" it is possible to set the time
after which the apnoea back-up function to support patient
ventilation, will be activated. The duration of the apnoea time
can be set within 5 and 60 seconds.
Da 5 a 60 s.
It is recommended to pay much attention to such parameter
setting because an incorrect adjustments could lead to serious
consequences for the patient: in particular we suggest to set a
value around 20 seconds.
It is possible to select the graphs displaying modality:
Single Charts in the graphs displaying area, it is displayed a
sole graph.
Double Charts in the graphs displaying area, two graphs are
displayed.
User's Manual, version DU3300105