6.5.3 Preliminary tests - Lung ventilator

The choice of the necessary entries to activate the preliminary tests on lung
ventilator, is made by setting the parameters and selecting the boxes of the
respiratory physiological parameters (hereafter RPP).
VC/VAC Operative mode
Set the following respiratory physiological parameters.
Vti
RR
I/E
Trig. I
PEEP
Pausa
MORPHEUS
For physiological respiratory parameters checking, see cfr. 5.2
and/or 7.
The VC/VAC is a volume controlled ventilation (Vti), synchronized
with thel patient breaths if the inspiratory trigger is activated (Trig. I).
500 ml
12 bpm
1:2
-6
5
10
For tests and checks the suggested patient simulator is SIARE code
LS.AB.001 which is equipped with variable compliance and
resistance.