6.3.8 Use of antibacterial filter

Apply the antibacterial filters to the patient circuit.
WARNING! Risk of injury for the patient.
Replace the antibacterial filters as per maintenance instructions (cfr.
10).
WARNING! Risk of injury for the patient.
To protect the patient from particles and dust, generated for example by
the soda lime, it is necessary to use a filter between the inpiratory hose
and the patient, i.e., the filter on the Y connector or the filter on
inspiratory hose.

6.3.9 Gas analyzer connection

Apply to the patient circuit the gas analysis
modules at the connector for services (24) as
shown in the picture.
Extract the gas analyzer module, provided
of the special cable, from package.
By the supplied "interface cable", connect
the gas analyzer connector to the 9 poles
connector positioned on the rear side of the
anaesthesia unit.
IRMA: apply the gas analyzer module, on
the patient circuit Y connector.
ISA: apply the sampling line on the patient
circuit Y connector.
For futher information on operating logic of GAS SENSOR
function,make reference to the relevant paragraph.