Siare Morpheus E User Manual page 253

Anaesthesia unit
Table of Contents

Advertisement

Initialization phase
Keys and encoder
knob
Error in the bus
"CAN"
Patient circuit
disconnected
9-2
The initialization phase is not completed and the system is
blocked.
Verify and intervene in function on the error messages and
indications evidenced during the "SELF TEST" phase.
Turn off and on the anaesthesia unit and repeat "SELF TEST"
phase.
Contact the Siare Service Centre or a Centre authorised by Siare.
The keys or the encoder knob do not work
Switch the anaesthesia unit OFF and then switch back ON.
Contact the Siare Service Centre or a Centre authorised by Siare.
This alarm condition occurs in case of system failure (electronic
boards).
Contact the Siare Service Centre or a Centre authorised by Siare
This alarm conditions occurs in case of pneumatic circuit
malfunctioning.
Check that the alarm limits are set correctly.
Check that the mask, endotracheal tube and patient circuit are not
in some way split, disconnected or connected wrongly If this is the
case, eliminate the problem or replace them.
Check the correct settings of the patient's respiratory parameters
(according to the operative mode selected: Volume/Flow, Rate,
I/E, Trigger).
Check that the patient circuit is connected correctly to the
anaesthesia unit and to the patient.
If this is not the case, contact the Siare Service Centre or a Centre
authorised by Siare.
User's Manual, version DU3300105