Rate low
Apnoea
MORPHEUS
This alarm is activated when the breathing rate volume is lower
than the set value.
•
Check that the corresponding alarm limits are set correctly.
•
Check that the patient's respiratory parameters are set correctly.
•
Check that the anaesthesia unit operates correctly, checking the
airways pressure trend. If the anaesthesia unit operates correctly,
check the flow sensor and the correct connection of its cable.
•
Check that the mask, endotracheal tube and patient circuit are not
in some way split, disconnected or connected wrongly if this is the
case, eliminate the problem or replace them.
•
Check that the sensitivity of the Trigger is appropriate to the
patient's physiological conditions.
•
Check that the patient circuit is connected correctly to the
anaesthesia unit and to the patient.
•
Contact the Siare Service Centre or a Centre authorised by Siare
In this condition, no spontaneous respiratory activities is
detected (RR = 0).
•
Check that the patient's respiratory parameters are set correctly.
•
Check that the sensitivity of the Trigger is appropriate to the
patient's physiological conditions.
•
Check that the mask, endotracheal tube and patient circuit are not
in some way split, disconnected or connected wrongly If this is the
case, eliminate the problem or replace them.
•
Check that the anaesthesia unit operates correctly, checking the
airways pressure trend. If the anaesthesia unit operates correctly,
check the flow sensor and the correct connection of its cable.
•
Check that the patient circuit is connected correctly to the
anaesthesia unit and to the patient.
•
Contact the Siare Service Centre or a Centre authorised by Siare.