Rate low
Apnoea
MORPHEUS
This alarm is activated when the breathing rate volume is lower
than the set value.
Check that the corresponding alarm limits are set correctly.
Check that the patient's respiratory parameters are set correctly.
Check that the anaesthesia unit operates correctly, checking the
airways pressure trend. If the anaesthesia unit operates correctly,
check the flow sensor and the correct connection of its cable.
Check that the mask, endotracheal tube and patient circuit are not
in some way split, disconnected or connected wrongly if this is the
case, eliminate the problem or replace them.
Check that the sensitivity of the Trigger is appropriate to the
patient's physiological conditions.
Check that the patient circuit is connected correctly to the
anaesthesia unit and to the patient.
Contact the Siare Service Centre or a Centre authorised by Siare
In this condition, no spontaneous respiratory activities is
detected (RR = 0).
Check that the patient's respiratory parameters are set correctly.
Check that the sensitivity of the Trigger is appropriate to the
patient's physiological conditions.
Check that the mask, endotracheal tube and patient circuit are not
in some way split, disconnected or connected wrongly If this is the
case, eliminate the problem or replace them.
Check that the anaesthesia unit operates correctly, checking the
airways pressure trend. If the anaesthesia unit operates correctly,
check the flow sensor and the correct connection of its cable.
Check that the patient circuit is connected correctly to the
anaesthesia unit and to the patient.
Contact the Siare Service Centre or a Centre authorised by Siare.