Set the flow level (pressure) to recognize the patient spontaneous
activity. (L/min) - (cmH2O).
Set the airways positive pressures support value in inspiratory
phase.
Set the airways positive pressure value in expiratory phase.
Set the time establishing the minimum duration of inspiratory
period. Therefore, in case the inspiratory phase has a duration
lower than such set value, the patient will be forced to remain in
inspiratory phase for such time.
Set the back-up respiratory rate used when the apnoea condition
occurs to activate a ventilation mode controlled by the unit.
Set the flow level (pressure) to recognize the patient spontaneous
activity. (L/min) - (cmH2O).
Set the expiratory flow percentage respect to the max peak of the
inspirited flow when the inspiration stops.
Set the flow value during the inspiratory phase.
Set the time establishing the duration of the inspiratory period of
the ventilator. It is possible to set values in function of RR
parameter set.
Set the value of guaranteed tidal volume at each breathing act.
5-35