4.3.2 Electronic flowmeter box alarm

The electronic flowmeter box of
MORPHEUS anaesthesia unit is
provided with a display area for
alarms view (4.3) and a membrane
key for their silencing (4.2).
The alarms operating logic - set values from flowmeter box menu.
LOW O
2
LOW N
O PRESSURE
2
LOW XENON PRESSURE
LOW AIR PRESSURE
FRESH GAS OCCLUSION
MORPHEUS
PRESSURE
Alarm detecting the oxygen failure or low O2
pressure.
Block of nitrous oxide delivery: CUT-OFF
alarm.
Alarm detecting the N 2 O failure or low N 2 O
pressure.
Alarm detecting the Xe failure or low Xe
pressure.
Alarm detecting the Air failure or low Air
pressure.
Alarm detecting the an overpressure and
enable the protection of flowmeter box
components.