Siare Morpheus E User Manual page 245

Anaesthesia unit
Table of Contents

Advertisement

Min. Exp. Volume
Max. Volume Exp.
FiO 2 Min.
FiO 2 Max.
8-24
This alarm signals the overcoming of the low limit of the
corresponding parameter. This is a high priority alarm,
activated after a delay of 15 s.
An acoustic indicator and a flashing string in the alarms zone,
as visual signal, are provided. The sound can be silenced and
suspended for 30 s; the inhibition is not foreseen.
This alarm signals the overcoming of the high limit of the
corresponding parameter. This is a high priority alarm,
activated after a delay of 15 s.
An acoustic indicator and a flashing string in the alarms zone,
as visual signal, are provided. The sound can be silenced and
suspended for 30 s; the inhibition is not foreseen.
This alarm signals the overcoming of the low limit of the
corresponding parameter. This is a high priority alarm,
activated after a delay of 15 s.
An acoustic indicator and a flashing string in the alarms zone,
as visual signal, are provided. The sound can be silenced and
suspended for 30 s; the inhibition is not foreseen.
If you set a FiO2 at 21%, the alarm is automatically disabled.
This High priority alarm, can be enabled even if the system
detects the cell disconnection. The value of the cell
disconnected bring to a displaying of the 0 value on the
display in the box of the O2 %.
This alarm signals the overcoming of the low limit of the
corresponding parameter. This is a high priority alarm,
activated after a delay of 15 s.
An acoustic indicator and a flashing string in the alarms zone,
as visual signal, are provided. The sound can be silenced and
suspended for 30 s; the inhibition is not foreseen.
If you set FiO2 at 21%, the alarm is automatically disabled.
User's Manual, version DU3300105