8.4
List of alarms and priorities
8.4.1 Lung Ventilator – configurable alarms
Alarm type
High Pressure
Low Pressure
High Frequency
Low Frequency
Min. Exp. Volume
Max. Volume Exp.
FiO 2 Min.
FiO 2 Max.
Power Supply
Fault
MORPHEUS
Priority
Delay time (s)
3 patient
HIGH
breaths
HIGH
15
3 patient
HIGH
breaths
HIGH
15
HIGH
0
HIGH
0
HIGH
15
HIGH
15
HIGH
0
Suspendable
Delay (s)
YES
30
YES
30
YES
30
YES
30
YES
30
YES
30
YES
30
YES
30
YES
120
Inhibition
NO
NO
NO
NO
NO
NO
YES
(auto if
FiO2<21%)
YES
(auto if
FiO2<21%)
YES
8-11