6.3.10 Mains power supply

The electrical connections are a very important part in the installation of the
anaesthesia unit. Incorrect connections or connections to unsuitable electrical
systems can compromise the safety of the patient and the operator.
The mains power supply must comply with the CEI 64-8/7 regulations concerning
type A premises for medical use.
The power supplies foreseen on MORPHEUS anaesthesia unit are of two types:
by main power supply;
by internal battery.
Connection to the mains power supply
The main electric power supply must correspond
to that indicated on the identification label (electric
power supply, frequency and power consumption)
located on the back side of anaesthesia unit:
100 ÷ 240Vac / 47 ÷ 63Hz / 200W
Connect the power cable outlet to the plug on
the anaesthesia unit.
Insert the plug of the power cable in the main
power supply outlet.
Position the main switch (located in the back
side of the anaesthesia unit) in position "I".
Verify on ventilator module front side (user's
control area) if the green led is on (it indicates
the presence of main power supply).
WARNING! Risk of personal-physical injuries
To prevent the risk of electric shock, connect the power cable of the
lung ventilator to an earthed socket.
6-16