Adjust the O 2 flow.
Verify that the sum of flows set on the display
of flowmeter box corresponds to the foreseen
value.
Adjust the concentration of anaesthetic agent
onto the vaporiser.
On the control panel for manual ventilation,
select the type of desired breathing circuit:
AUX: control of fresh gas exit selection
APL : enabling of valves group
MORPHEUS
The anaesthesia unit is ready to start the phase of patient ventilation
according with the modalities described in the following paragraphs.
Electric control.
The activation allows to enrich
the mixture of gas delivered to
the patient, conveying pure
oxygen (flow of around 35 l/min.)
in the valves group or in the TO
and FRO patient circuit with the
aim of cleaning the circuit in case
of emergency, early awakening,
etc.