Siare Morpheus E User Manual page 161

Anaesthesia unit
Table of Contents

Advertisement

LEAK TEST
Before performing the Leak Test, ensure that:
6-36
WARNING! - Risk of anaesthesia unit failure and/or injuries for
the patient.
This test verifies that there are no leaks higher than 100 ml in the
anaesthesia unit pneumatic circuits.
WARNING! - Risk of injury for the patient.
Perform the Leak Test weekly.
Select the Leak Test entry.
Press the encoder: the Leak Test is activated.
the gas flow regulators of flowmeter box are closed;
the CO 2 absorber canister is not mounted on valves group;
the supplied manual ventilation kit is not connected to the valves
group connector.
User's Manual, version DU3300105