9
TROUBLESHOOTING
This chapter is a guide for the operator and the technician, providing indications for
eliminating, as quickly as possible, most of the problems that may have caused
malfunctioning or alarm signals.
This chapter describes the possible causes of problems, indicated by alarms that are
activated during normal functioning.
9.1
List of malfunctioning
No power
Power supply
MORPHEUS
If the problem persists, carry out a complete check of the
MORPHEUS anaesthesia unit to identify any irregularities.
If the problem cannot be resolved, contact the Siare Service Centre
or a Centre authorised by Siare.
The anaesthesia unit does not switch on
•
Check that it is connected to the main power supply
•
Check that the main switch is turned to the I position (ON)
•
Check the main fuses
•
Contact the Siare Service Centre or a Centre authorised by Siare.
There is a power supply fault and the anaesthesia unit is
operating on the battery
•
Check that it is connected to the main power supply
•
Check that the main switch is turned to the I position (ON)
•
Check the correct connections of the plug, the fuses and the
connector, and the condition of the cable (if necessary, restore the
connections and replace the cable if it is damaged).
•
Check that power is present at the relative socket by plugging in
another electrical device. If there is no power, use another socket
or check the overload switch on the electrical panel of the room.
9-1