GAS SENSOR - ZERO CALIBRATION
•
•
MORPHEUS
WARNING – Risk of failure of anaesthesia unit and/or injury to
the patient.
This PROCEDURE, to be used to perfom the zeroing of gas
analyzer, is ENABLED only when a GAS PROBE is connected to
the relevant connector positioned on the back side of the unit for
anaesthesia.
It is possible to perform the Zero Calibration of the gas probes also
by the controls: MENU' – SETUP – FOLLOWING PAGE – GAS
PROBE.
Before performing the Zero Calibration, ensure that the electric
connecting cable of the probe is correctly connected.
•
ISA sidestream probe : before performing the Zero
Calibration, consult instructions to relative User's Manual.
•
IRMA mainstream probe : before performing the Zero
Calibration, consult instructions to relative User's Manual.
Select the item, Gas Probe: Zero Calibration
Press the encoder: the Gaz Probe zeroing procedure is activated.
6-39