Low pressure
High pressure
High/low oxygen
concentration
Power failure
Gas failure
6-50
•
Set thelow pressure alarm limit at 5 cmH 2 O.
•
During anaesthesia unit operation discounnect the patient
simulator from patient circuit (SIARE type).
•
After around 15 seconds the low airways pressure alarm
occurs: silence the alarm.
•
Re-connect the patient simulator.
•
Set the high pressure alarm limit at 45 cmH 2 O.
•
During the anaesthesia unit operation block (wit the hands)
the patient simulator.
•
The high airways pressure alarm occurs: silence the alarm.
•
Unlock the patient simulator.
See cfr. 6.5.2
•
During the anaesthesia unit operation place on OFF ( 0 )
the main switch (cfr. 6.3.10).
•
The power failure alarm occurs: silence the alarm.
•
Replace the main switch in ON position ( I ).
•
During anaesthesia unit operation close the medical gas
supply
•
The gas supplì failure alarm occurs: silence the alarm.
•
Restore the medical gas supply.
WARNING! ! Risk of injuries for the patient
The alarms must be activated in the correct times and modes.
Verify the correct activation of the visual and acoustic indications.
User's Manual, version DU3300105