5.13.9 CALIBRATION PROGRAMS activation

To enable the CALIBRATION PROGRAM displaying the anaesthesia unit Morpheus
must be correctly connected and work.
Prepared for the use (cfr.: Maintenance, Cleaning and Disinfection and
Sterilization).
Correctly positioned.
All the accessories and devices for the functioning to be prepared.
Power supply and gases properly connected.
A patient simulator shall be connected to the patient circuit equipped with the
machine.
5-52
Anaesthesia unit Morpheus powered (ventilator switched off).
Press at the same time the buttons : STAND BY and
ALARM RESET.
After few seconds, the anaesthesia unit Morpheus
switches on and displays an area where the programs
needed for the calibration are available.
See Service Manual for further details and instructions on the use of
CALIBRATION PROGRAMS.