Low gas pressure
Battery charge level
25% (50%)
Low battery (10
minutes)
Exhausted O2
sensor
MORPHEUS
This alarm is activated when the pressure is insufficient (< 2.7
bar) for the anaesthesia unit to operate correctly.
Check that the medical gases are correctly connected to the
anaesthesia unit. Restore the connections or replace the tubes if
damaged.
Check that there is sufficient pressure in the supply system or in
the cylinders. Adjust or repair the supply system (or replace the
cylinders) if the pressure is insufficient.
Contact the Siare Service Centre or a Centre authorised by Siare.
This alarm is activated when the charge level of the battery is at
25% (50%) of the fully charged level: no more than 30 (60)
minutes of operating autonomy is guaranteed.
Check that it is connected to the main power supply.
Recharge the battery.
If the alarm is activated when the battery has not provided the
time autonomy indicated on the technical sheet, request the
intervention of a Service Centre.
This alarm condition is present when the charge battery level is
such to be guaranteed a residual autonomy of about 10
minutes.
Verify the correct connection of power supply.
To recharge the battery.
If the alarm is activated when the battery has not provided the
time autonomy indicated on the technical sheet, request the
intervention of a Service Centre.
Low O 2 concentration
reagents in the oxygen sensor are exhausted.
Replace the oxygen sensor with a new one.
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.
This alarm is activated when the
-
9-3