Low gas pressure
Battery charge level
25% (50%)
Low battery (10
minutes)
Exhausted O2
sensor
MORPHEUS
This alarm is activated when the pressure is insufficient (< 2.7
bar) for the anaesthesia unit to operate correctly.
•
Check that the medical gases are correctly connected to the
anaesthesia unit. Restore the connections or replace the tubes if
damaged.
•
Check that there is sufficient pressure in the supply system or in
the cylinders. Adjust or repair the supply system (or replace the
cylinders) if the pressure is insufficient.
•
Contact the Siare Service Centre or a Centre authorised by Siare.
This alarm is activated when the charge level of the battery is at
25% (50%) of the fully charged level: no more than 30 (60)
minutes of operating autonomy is guaranteed.
•
Check that it is connected to the main power supply.
•
Recharge the battery.
•
If the alarm is activated when the battery has not provided the
time autonomy indicated on the technical sheet, request the
intervention of a Service Centre.
This alarm condition is present when the charge battery level is
such to be guaranteed a residual autonomy of about 10
minutes.
•
Verify the correct connection of power supply.
•
To recharge the battery.
•
If the alarm is activated when the battery has not provided the
time autonomy indicated on the technical sheet, request the
intervention of a Service Centre.
Low O 2 concentration
reagents in the oxygen sensor are exhausted.
•
Replace the oxygen sensor with a new one.
•
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.
This alarm is activated when the
-
9-3