6.2
Before the use
6.2.1 Assembling of O 2 cell
•
Unpack carefully the O 2 cell
•
Lift the valves group cover.
•
Insert and screw the cell, in the space
provided with the "O2 SENSOR" script.
•
Ensure that the electric cable for cell
connection is positioned in the
dedicated space.
•
Connect the plug on the O 2 sensor.
•
Connect the RJ connector on the
dedicated intake within the valves
group.
WARNING! Risk of injury for the patient
Check of electric connection of O 2 cell.
At each anaesthesia unit start-up, the system check the presence of
the electric connection to the O 2 cell.
"SELF TEST" phase " (cfr. 6.4.1) or cfr. 7.
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
WARNING. Danger of injuries to the user
To avoid the danger of electric shocks and/or break of component
parts, during interventions, ensure that the electric power supply
disconnected from anaesthesia unit.
Preparation to use
6-5