Medical gas supply
Electric power supply
Back-up battery
Alarms
Safety devices
4-4
OXYGEN
Pressure included between 280 kPa and 600 kPa (2,8 - 6 bar)
Max. required flow 90 l/min.
NITROUS OXIDE
Pressure included between 280 kPa and 600 kPa (2,8 - 6 bar)
Max. required flow 15 l/min.
MEDICAL COMPRESSED AIR
Pressure included between 280 kPa and 600 kPa (2,8 - 6 bar)
Max. required flow 90 l/min.
100 ÷ 240Vac / 45 ÷ 60Hz
12Vdc - 3,0 Ah Pb battery; it grants 120 minutes operation
Lack or low oxygen pressure with consequent cut-off of
nitrous oxide delivery
Lack or low nitrous oxide pressure
Lack or low xenon pressure
Lack or low air pressure
Fresh gas occlusion
AGAINST THE ADMINISTRATION OF HYPOXIC MIXTURES
MIX-LIFE: it always guarantees a minimum concentration of 25 %
oxygen on mixtures which includes nitrous oxide.
IN CASE OF LACK OR LOW OXYGEN PRESSURE CUT-OFF:
audible alarm with immediate cut-off of nitrous oxide delivery.
AGAINST FRESH GAS HIGH PRESSURE IN SUPPLY: pressure
sensor for protection of flowmeter box components.
IN CASE OF LACK OR COMPRESSED AIR LOW PRESSURE :
all the devices (gas feeding) supplied by compressed air are
automatically supplied by oxygen.
AGAINST THE SIMULTANEOUS DELIVERY OF AIR AND N
selection keyboard on the flowmeter front panel.