•
Open the O 2 (4.11) regulator or AIR (4.6) at
2 l/min.
•
Press the AUT key to activate the VC/VAC
ventilation (cfr. 6.5.3)
•
The anaesthesia unit begins the inspiratory
(expiratory) phase determining a variation of
parametes measured in the monitoring area
of ventilator.
•
Check the regular operation of the
anaethesia unit.
O 2 concentration
•
Open the O 2 regulator (4.11) at 7 l/min.
•
Verify that the oxygen concentration of
parameters measured in the monitoring area
of ventilator [ O 2 ], increases.
•
Close the O 2 regulator.
6-46
For setting the respiratory physiological parameters (hereinafter
RPP) useful for preliminary tests on lung ventilator, consult the cfr.
6.5.3.
Verify the presence of soda lime in the jar and that it is not
exhausted.
User's Manual, version DU3300105