Open the O 2 (4.11) regulator or AIR (4.6) at
2 l/min.
Press the AUT key to activate the VC/VAC
ventilation (cfr. 6.5.3)
The anaesthesia unit begins the inspiratory
(expiratory) phase determining a variation of
parametes measured in the monitoring area
of ventilator.
Check the regular operation of the
anaethesia unit.
O 2 concentration
Open the O 2 regulator (4.11) at 7 l/min.
Verify that the oxygen concentration of
parameters measured in the monitoring area
of ventilator [ O 2 ], increases.
Close the O 2 regulator.
6-46
For setting the respiratory physiological parameters (hereinafter
RPP) useful for preliminary tests on lung ventilator, consult the cfr.
6.5.3.