PRESSURE SUPPORT
With this parameter it is possible to set the value of the positive pressure of
airways in inspiratory phase.
It is possible to adjust this parameter on values from 6 to 60 cmH
steps of 1 cmH
O
2
This parameter is used in the following operative modalities: PSV, SIMV.
RESPIRATORY RATE
With this parameter it is possible to set the number of breaths to be
effected by the patient in one minute.
The set values are from 4 to 120 bpm with step of 1. In APCV
pressometric modality, the set RR values vary upon the set Ti.
With Ti = 0.5 s from 24 to 50 bpm
With Ti = 1.0 s from 12 to 48 bpm
With Ti = 1.5 s from 08 to 32 bpm
With Ti = 2.0 s from 06 to 24 bpm
With Ti = 2.5 s from 05 to 19 bpm
With Ti = 3.0 s from 05 to 16 bpm
BACKUP RESPIRATORY RATE
With this parameter it is possible to set the back up respiratory rate in PSV,
used when the APNOEA condition occurs to activate a ventilation modality
controlled by the ventilator.
The values can be set from 5 to 50 bpm with steps of 1 in the pressometric
modalities (PSV) according with Ti set.
PSV
With Ti = 0.5 s from 24 to 50 bpm
With Ti = 1.0 s from 12 to 48 bpm
With Ti = 1.5 s from 08 to 32 bpm
With Ti = 2.0 s from 06 to 24 bpm
With Ti = 2.5 s from 05 to 19 bpm
With Ti = 3.0 s from 05 to 16 bpm
O with
2
7-33