Couplings and
Dismantle and clean, then sterilize in an
connectors
autoclave, disinfect with steam or chemically.
Dismantle and clean, then sterilize in an
Breathing system
autoclave, disinfect with steam or chemically.
Flow sensor
Disinfect with steam or chemically.
MORPHEUS
Do not clean or re-use disposable circuit tubes.
Do not clean or re-use if the filters are the disposable
type.
WARNING! Risk of injury for the patient
It is necessary to have at least one spare patient circuit
in stock for routine use and /or accidental breaks.
WARNING. Risk of device failure.
Do not attempt to dismantle or clean with compressed
air.
The flow sensor can be washed and disinfected by
immersing it in a bowl with 3 centimetres of liquid,
keeping the connector for the electrical connections
facing upwards.
Check that there are no
splits and replace them if
they are damaged.
Before using again,
eliminate any humidity
inside the components by
means of compressed air.
Check that there are no
splits and replace them if
they are damaged.
Before using again,
eliminate any humidity
inside the components by
means of compressed air.