Control panel for manual ventilation
On the left side of work-shelf it is positioned a small control panel to
manage all the operations of manual ventilation with the valves group.
This panel includes the following functions.
The APL valve for the adjustment of airways maximum pressure
during manual ventilation.
The pressure increases by turning clockwise the valve knob and
it decreases by turning counter-clockwise.
The range of regulation is around from 0 to 50 cm H 2 O.
Electronic O 2 BY-PASS button. By pressing this button pure
OXYGEN is released into the anaesthesia circuit with a flow of
approximately 35 l/min.
Fresh gases exit selection control: enabling of fresh gases exit
connector (AUX).
Fresh gases exit selection control: enabling of valves group
(APL).
3-14
Thanks to this type of control it will be possible to enable the
REMOTE option, a remote control that could be positioned by the
doctor near the patient with the aim to facilitate the operativity of
induction and awakening manoeuvres.