8.4.2 System alarms

Alarm type
APNOEA
BATTERY LOW 50%
REM.
BATTERY LOW 25%
REM.
LOW BATTERY 10
MIN. REM.
TURBINE FAILURE
TURBINE
OVERTEMPERATURE
TURBINE
OVERCURRENT
LOW GAS SUPPLY
DISCONNECTED
CIRCUIT
CAN-BUS FAILURE
8-12
Priority
Delay time (s)
HIGH
MEDIUM
HIGH
HIGH
HIGH
HIGH
HIGH
HIGH
MEDIUM
HIGH
The following series of SYSTEM STATE INDICATIONS are displayed
in the EVENTS
POWER ON
STAND BY
VENTILATION START
POWER OFF
Suspendable
0
YES
0
YES
0
YES
0
NO
0
YES
0
NO
0
YES
5
YES
0
YES
0
NO
User's Manual, version DU3300105
Delay (s)
Inhibition
30
NO
120
YES
60
YES
-
NO
30
NO
-
NO
30
NO
YES
120
(auto if
FiO2<21%)
120
YES
-
NO