Primary test: at anaesthesia unit's start-up, a control test of Turbine
presence, Medical Gas Supply presence, INSP and EXP flow sensors
operation, pressure sensor, patient circuit losses, back-up battery state,
oxygen cell, integrity of audible alarm is automatically performed. This test
takes around 15 seconds.
Tests on demand: the anaesthesia unit has a tests on demand which is
activated by the user in the ventilator menu. This subtest permits to verify the
dead space and losses or to perform the oxygen cell calibration.
APCV
APCV-TV
PSV
APNEA BACK-UP
SIMV (Volumetric +PS; SPONT)
VC/VAC
VC/VAC BABY (integrated NEONATAL ventilation mode)
MANUAL
From 4 to 120 bpm (step 1 bpm)
1:1, 1:1.5, 1:2, 1:3, 2:1, 3:1
From 0,2 to 5 sec.
From 0 to 60% of inspiratory time
From 1 ÷ 119 bpm
From 5 to 1500 ml (< 50ml: step 1ml / 50-100ml: step 5ml / >100ml step 10ml)
From 1 to 30 liters
From 6 ÷ 60 cmH 2 O
From 5 ÷ 60 cmH 2 O
OFF, 3 ÷ 30 cmH 2 O (step 1 cmH 2 O)
Selectable ramp from 10 to 80 L/min (step 10 L/min)
Minimum resolution 1% - Automatic calibration procedure
-20 ÷ 80 cmH 2 O
User's Manual, version DU3300105