The MORPHEUS anaesthesia unit complies with the requirements on
electromedical equipment stated in the IEC/EN 60601-1-1 and IEC/EN
60601-1-2 directive.
To ensure correct functioning of the MORPHEUS anaesthesia unit, it is
advisable that any equipment to which it is connected also complies
with the regulations mentioned above.
Battery power supply
To silence the acoustic alarm, push the
key " ALARM RESET "; located on the
ventilator front panel.
When the led on the front panel is on it indicates that
the anaesthesia unit is correctly supplied.
The operation time of the anaesthesia unit guaranteed by the battery
can vary in the following cases:
MORPHEUS
The battery, inside the MORPHEUS anaesthesia unit, must always
be installed.
If the battery is not fitted, the anaesthesia unit is not protected
against fluctuations or failures of the power supply.
The anaesthesia unit must not be used without a mains power
supply.
The use of the battery must be limited to short periods and is not
foreseen as an alternative to the mains power supply.
Do not open the mains power supply module to replace the battery
or to carry out maintenance operations on the battery charger.
old battery or not perfectly efficient
not standard ventilatory parameters
presence of electronic flowmeter box.