4.2

Before the use

4.2.1 O
cell assembling
2
Insert and screw the cell: in the space marked with
the script " O2 CELL ".
Connect the electric cable of the cell passing it into
dedicated space.
The O
The RJ connector on the specific socket within
the valves group.
WARNING !! Risk for Patient injury
At each anaesthesia unit start-up, the system checks the presence of the electric
connection to the O

4.2.2 Assembling of absorber canister

On the anaesthesia unit it is possible to use a multipurpose rechargeable absorber
canister or a disposable pre-loaded absorber canister.
It is recommended to use only original SIARE spare parts or spare parts checked
and approved by SIARE.
The absorber canister is positioned in the upper side of
valves group. The specific lock (unlock) lever makes
very easy to hook (unhook) the canister.
Mount the CO
2
unit as shown in the picture.
To unpack carefully the CO
Shake the absorber canister (disposable model) in
order to separate the soda lime granules.
Remove the seal from absorber canister
(disposable model).
Insert the absorber canister in the apposite groove
(opposite side of the "PUSH" lever).
Morpheus_ND
cell connector
2
Visual check of electric connection of oxygen cell.
cell ( Self Test phase, see on chapter 5.2 or chapter 4.7 )
2
absorber canister on the anaesthesia
absorber canister.
2
4-3