Main Menu - Trends - Siare Morpheus E User Manual

Anaesthesia unit
Table of Contents

Advertisement

5.9.3 Main Menu – TRENDS
The system through the TRENDS modality, enable the user to monitor the most
significant breathing parameters on medium – long period of a patient. The
examined PRF, monitored and displayed in sequences and in pairs are:
the respiratory rate or breaths of the patient - RR (bpm)
the airways pressure - PAW (cmH2O)
the end expiration positive pressure - PEEP (cmH2O)
the minute volume expired by the patient - VM (L)
the expired tidal volume - Vte (ml)
MORPHEUS
Rotate the encoder to select the desired item of MAIN MENU area
[ e.g. TRENDS ].
Press the encoder to display the page of desired item of MAIN
MENU [ e.g. TRENDS ].
The capacity to store each parameter is of 72 hours with data
sampling at 4 minutes.
5-23