8.1

Definitions

Disable an alarm
Alarm inhibition
Alarm restore
Alarm stop and
silencing
Alarm suspension
Alarm signal
High priority alarm
signal; warning
Medium priority
signal;
precautionary
Low priority signal;
warning
Information signal.
8-2
Inactivate an alarm function.
Abolition of alarm signals which persists unless intentionally
revocated.
Make that the alarm function is restored to its default initial
state.
Causing the cessation and restore of an acoustic alarm by a
deliberate action.
Abolition of alarm signals which persists unless automatically
cancelled
It indicates an occurring condition and/or the duration of a
condition which requires a reply by the user.
It indicates the need of an immediate reply by the user.
It indicates the need of a prompt reply by the user.
It indicates the need of attention by the user.
An acoustic visual signal or a combination of both, whose aim
is to supply technical or physiological information.
The ventilator module is in accordance to what is provided by the EN
475 standard (Electrically-generated alarm signals) and EN ISO
9703-3 standard (Anaesthetic and respiratory care alarm signals.
Guidance on application of alarms).
User's Manual, version DU3300105