6.3.5 Patient circuit connections

Insert the patient circuit supporting arm in the specific bracket, fixed to the full-height
steel rod used for accessories mounting.
Connect the supplied patient circuit to the
specific INS and EXP connectors on the
valves group.
Position the patient circuit on the patient
circuit supporting arm.
6-12
Use a patient circuit
suitable for the patient to
ventilate.
WARNING! Risk of injury for the patient.
Whenever the circuit is changed or replaced, it is necessary to
perform the LEAK TEST to verify eventual leaks and the compliance
of patient circuit (Preliminary tests cfr. 6.4).
The system checks the patient circuit at every start-up of the unit:
"SELF TEST" phase (Preliminary tests cfr. 6.4 or cfr. 7).
WARNING. Risk of strangling.
Pay special WARNING! when connecting the patient circuit to the
anaesthesia unit. If not carefully positioned, the hoses, the cables,
the patient circuit and other similar components to the anaesthesia
unit, these can be dangerous for the patient.
WARNING. Risk of burns
Do not use conductive masks or conductive breathing hoses during
surgery with electrosurgical units because they can cause burns.
Tidal Volume
< 50 mL
from 50 to 200 mL
> 210 mL
User's Manual, version DU3300105
Set of hoses
Neonatal
paediatrics
adults