6.3.12 Connection to other equipments

Connection to Siare equipment
If the equipment to be connected is a SIARE unit, all the instructions
necessary for the connection to anaesthesia unit can be found in the
documentation supplied with the unit.
WARNING! Risk of injury for the user / patient
Do not connect external devices NOT manufactured or NOT authorized
by SIARE to the anaesthesia unit (e.g., scavenging systems, patient
simulators), and not described in the present user's manual.
In case of need contact SIARE.
WARNING! Risk of injuries for th patient
When using additional components in the respiratory systems or
configurations not conform to those supplied with the anaesthesia unit,
the inspiratory and expiratory resistence can increase until overcome
the standard requirements.
When using configurations of this type, it is necessary to pay particular
WARNING! to the measuring values.
WARNING! Risk of electric shock.
In case of malfunctioning of ground conductor, the connection of other
electric equipment to the supplementary outlets of the anaesthesia unit
could cause an increase in leakage current over the values allowed by
the law.
In case of connection of other devices to the supplementary outlets, it is
necessari to verify the total leakage current.
In case of overcoming of the values allowed for total leakage current, do
not connnect the devices to the supplementary outlets of anaesthesia
unit, but to separate electric outlets.
All the system must satisfy the requirements for electromedical
equipment stated by the IEC/EN 60601-1-1 and IEC/EN 60601-1-2.
directives.
6-20
User's Manual, version DU3300105