Disconnected O 2
cell
FiO 2 high
low
FiO 2
9-4
Low O 2 concentration
status of the oxygen sensor.
Check that the oxygen cell is correctly connected.
Replace the oxygen sensor with a new one.
Check the condition of the cable and the connector (if necessary,
restore the connection and replace the cable if damaged).
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.
This alarm is activated when the measured concentration of
oxygen exceeds the set limit.
Check that the corresponding alarm limits are set correctly.
Calibrate the oxygen cell: if the problem occurs again after a short
time, replace the oxygen cell.
Contact the Siare Service Centre or a Centre authorised by Siare.
This alarm is activated when the measured concentration of
oxygen is below the set limit.
Check that the oxygen cell is fitted correctly in its housing.
Check that the corresponding alarm limits are set correctly.
Calibrate the oxygen cell: if the problem occurs again after a short
time, replace the oxygen cell.
Check that the feeding pressure of the medical gases is correct: if
it is not, check the pressure of the distribution system and the
correct connection to the supply.
Check that the mask, endotracheal tube and patient circuit are not
in some way clogged, bent or crushed. If this is the case, eliminate
the problem or replace them.
Contact the Siare Service Centre or a Centre authorised by Siare.