Ti
Ti bk
Ti min
Vte
Vti
MORPHEUS
INSPIRATION TIME
With this parameter it is possible to set the time establishing the duration of
the ventilator inspiration period.
It is possible to select values from 0.2 a 5.0 sec. sec. with steps of 0.1
seconds according with RR parameter set.
This parameter is combined with APCV, VC/VAC BABY, and SIMV
operative modes.
BACKUP INSPIRATORY TIME
With this parameter it is possible to set the back-up time in PSV, used in
case of APNOEA condition to activate a ventilation modality controlled by
the ventilator.
It is possible to select values from 0.2 to 5.0 sec. with step of 0.1 seconds
in function of set RRbk parameter.
MINIMUM INSPIRATORY TIME
With this parameter it is possible to set the time establishing the minimum
duration of inspiratory period. Therefore, in case the inspiratory phase has
a duration lower than such set value, the patient will be forced to remain in
the inspiratory phase for that time.
Through this parameter it is possible to set the minimum duration of
inspiratory phase of the assisted spontaneous breaths.
It is possible to select values from 0.3 to 2.5 sec. with steps of 0.1 sec.
This parameter is combined with the PSV modalities.
TIDAL VOLUME
With this parameter it is possible to set the quantity of gas mixture volume
guaranteed at each breath.
This parameter is used in the foreseen operative modalities: APCV-TV and
VA/VAC BABY.
It is possible to select values from: 0.005 a 1.0 L. (<50ml: step 1ml / 50-
100ml: step 5ml / >100ml step 10ml).
TIDAL VOLUME
With this parameter it is possible to set the quantity of gas mixture volume
guaranteed at each breath.
This parameter is used in the foreseen operative modalities: SIMV and
VA/VAC.
It is possible to select values from: 0.05 to 1.5 L. with steps of 10mL.