6.4.3 " SELF TEST " Verifications

Turbine
Gas Supply
Insp. Flow Sensor
Exp. Flow Sensor
Pressure Sensor
Patient Circuit
Battery
Oxygen Sensor
Acoustic Alarm
MORPHEUS
During the 'SELF TEST' phase, the ventilator software performs the
auto-diagnosis test and checks some devices necessary to the safe
functioning of the Morpheus anaesthesia unit.
Verification of SETUP – Turbine parameter (Enabled or
Disabled).
Verification of the presence of Air and O 2 supply pressure.
Verification of INSP flow sensor operation (with turbine option
only).
Verification of EXP flow sensor operation.
Verification of pressure sensor operation through control of
PAW reading.
Verification of patient circuit connection through checking of
pressure presence.
Checking on battery power.
Verification of O 2 cells connection in the valves group.
Verification by the user of acustic signal emission, the
confirmation of the test is made by silencing of that alarm.
For a correct and deeper analysis of the problems arising from the
"SELF TEST " phase, consult the service manual.
If the " SELF TEST " phase is not overcome consult the chapters 8
(Alarms) and 9 (Troubleshooting) or contact the nearest Siare
Service Centre or a center authorized by Siare.