6.5
Preliminary tests – Operating phase
The preliminary tests aer divided in four categories:
1. TEST ON DEMAND
O 2 Sensor Calibration
Leak Test
Gas Sensor: Zero Reference Calibration
2. Flowmeter module
3. Ventilator module
Respiratory parameters
Spirometry
4. Alarms
MORPHEUS
The anaesthesia unit must be ON, in STAND-BY operative mode,
once overcome the preliminary tests.
WARNING! - Risk of anaesthesia unit failure and/or injuries for
the patient.
A performance or cancellation of preliminary tests could determine a
malfunctioning during the unit operation: pay much warning.
Unless in case of emergency situation, always follow all preliminary
tests.
If the preliminary tests are not performed due to an emergency
situation, they could be performed in the more complete way as soon
as possible.