Parameters monitoring
Press the AUT key: the anaesthesia unit begins ventilating.
Basing on set respiratory physiological parameters and
on patient characteristics [patient simulator], the
ventilator module can measure and monitor a series of
significative magnitude for clinical evaluation.
Check the correspondence of measured parameters in
function of set ones.
The measured values are updated at each breath.
Vary the set parameters
values and verify their
correspondence on measured
parameters:
Vti - Vte
RR
I:E
VM
PEEP
FiO 2 (flowmeter box)
6-48
Ensure of the correct operation, checking also that:
the airways pressure increases during the inspiratory phase
the airways pressure limit intervenes (pressometric operative
mode)
the opening of oxygen flowmeter determines the increasing of
concentration on oximeter (O 2 %)
at varying of ventilation parameters, the anaesthesia unit
operation corresponds
the trigger operation is efficacious
the volume and respiratory rate set values are displayed
correctly and the pressure, volume and flow curves are in
accordance with monitored parameters.