High flows open circuit ( NON REBREATHING )
The breathing gas concentration is similar to the fresh gas ones.
There is a high consumption of fresh gas
The temperature and humidity of breathing gas are low. The high flow
ventilation is recommended only for pre-anaesthesia and awakening or for brief
surgeries.
There is not CO 2 accumulation in the patient circuit.
Semi-closed circuit ( REBREATHING )
The fresh gas flow in the patient circuit is lower than the ones breathed by the
patient; therefore the fresh gas concentration of the breathed gas is different
than the ones of the fresh gas, related to the BREATHED GAS / FRESH GAS
ratio
There is a medium consumption of soda lime.
There is a medium consumption of fresh gas.
The temperature and humidity of the fresh gas are acceptable for a long surgery
situation.
7-4
To perform this type of ventilation it is necessary to take off the soda
lime canister from valves group.
Required supervision: normal
Required supervision: pay attention to the gas concentration
monitoring (O
, N
O, CO
2
2
and halogenated gas).
2
User's Manual, version DU3300105