Battery Low 50%
rem.
Battery Low 25%
rem.
Low Battery 10
Min. rem.
Turbine Failure
8-26
This alarm is activated 0 sec. after the battery voltage falls
below 50% of the fully charged level.
It is a medium priority alarm indicated by an acoustic and visual
signal. It can be silenced.
Inhibiting the acoustic signal of this alarm is the same as that for
the power supply failure alarm (there is however a visual signal
in the alarm area).
This alarm is activated 0 sec. after the battery voltage falls
below 25% of the fully charged level.
It is a high priority alarm indicated by an acoustic and visual
signal. It can be silenced.
Inhibiting the acoustic signal of this alarm is the same as that for
the power supply failure alarm (there is however a visual signal
in the alarm area).
This is an alarm that is activated after a delay of 0 sec. by the
moment in which the residual autonomy of the battery is of
about 10 minutes.
Is an alarm of high priority, signalled by an acoustic and visual
device. He cannot be reset and even inhibited.
WARNING! Risk of injury for the unit
When this alarm signal is activated only 10
minutes of autonomy are available and the
machine will stop to ventilate.
This high priority alarm which is enabled after a 0 sec delay and
highlights the malfunctioning of the TURBINE.
WARNING! Risk of injury for the unit
Switch off immediately the device and get in
touch with the Technical service urgently.
User's Manual, version DU3300105