6.2.3 Battery charger

The anaesthesia unit is equipped with an internal battery that guarantees (if
perfectly efficient) at least 120 minutes operation (90 minutes, according to
ventilation parameters), in case of power failure.
The switching to battery operation is made automatically: on the ventilator screen
appears the relevant message "Power failure".
The anaesthesia unit battery can be recharged by connecting the equipment to
main power supply (using power cable supplied with the unit) and placing the
main switch in " I " position.
Carry a battery charge at least 10 hours, before using the the
anaesthesia unit the first time.
The operating time of anaesthesia unit guaranteed by the battery, can
vary in the following cases:
Follow the instructions.
Connect the outlet of the power cable to the
plug on the anaesthesia unit.
Insert the plug of the power cable in the
mains supply wall socket.
The electric main power supply must correspond
to the one indicated in the identification label on
the back side of the machine.
Turn the main switch (located on the back
side of the unit) in position "I".
6-6
old battery or not perfectly efficient
not standard ventilatory parameters
presence of electronic flowmeter box.
User's Manual, version DU3300105