Various Notes - Siare Morpheus E User Manual

Anaesthesia unit
Table of Contents

Advertisement

4.3.3 Various notes

4-16
WARNING! Risk of break of anaesthesia module.
To not compromise the duration and the correct operation of
regulators avoid to close them too energically.
The flow value must be read in correspondence of the upper edge of
flowmeter while it is rotating.
If the rotation is missing, consult the technical service.
WARNING! Risk of injury for the patient.
. If the anaesthesia module emits a shrill whistle
CUT-OFF ALARM
it means that the oxygen pressure is too low; it necessary to
immediately intervene to restore the line pressure.
If the line pressure is not available, open the emergency oxygen
cylinder.
Before using the anaesthesia unit on a patient it is necessary to per
form a series of preliminary checks to verify the correct equipment
operation.
Among the preliminary checks to be performed, it is also include the
flowmeter box.
User's Manual, version DU3300105