2. Enter the initial temperature.
a. Press METHOD to display Initial Temp.
b. Press UP or DOWN until you see 50 °C.
3. Enter the initial time.
a. Press NEXT to display Initial Time.
b. Press UP or DOWN to display initial time of 60 seconds.
4. Enter the ramp 1 rate.
a. Press NEXT to display Ramp 1 Rate.
b. Press UP or DOWN to set the Ramp Rate to MAX. This ramps the probe tip
ballistically at its maximum rate of approximately 150 °C/min.
5. Enter the ramp 1 temperature.
a. Press NEXT to display Ramp 1 Temp.
b. Press UP or DOWN until the temperature is 350 °C.
6. Enter the ramp 1 hold time.
a. Press NEXT to display Ramp 1 Hold.
b. Press UP or DOWN until the Ramp 1 Hold time displays 40 seconds.
7. Enter the ramp 2 rate.
a. Press NEXT to display Ramp 2 Rate.
b. Press UP or DOWN until the Ramp 2 Rate is OFF.
8. Return to DIP 1 Ready.
Press NEXT to display DIP 1 Ready.
9. Load the method.
Press PROG to display Load Method: X.
10. Save the method.
a. Press NEXT to display Save Method: X
b. Press ENTER to return to DIP 1 Ready.
c. Listen for the DPC to beep three times to indicate that your current method
configuration is saved and stored in the DPC as method 1, and may be recalled by
loading method 1.

To retrieve methods

1. Press the Prog key to display Load Method: X
2. Press Up or Down to locate the desired method number.