Setting Section Thickness and Trimming Thickness

The required section and trimming thicknesses are set
with the knob [A].
To choose between section thickness and
trimming thickness, press the knob [A]
In the NORMAL display mode the thickness range is indicated in the display.
FEED pre-selected section thickness
TRIM pre-selected trimming thickness
The graduation of the section thicknesses is divided into five ranges:
Range
up to 5 µm
from 5 µm to 20 µm
from 20 µm to 30 µm
from 30 µm to 60 µm
from 60 µm to 100 µm
24
Thermo Scientific Rotary Microtome HM 340E
Graduation
0,5 µm
1 µm
2 µm
5 µm
10 µm